For LAC release notes, see the following link – https://www.dmu-gmbh.at/en/products/releasenotes
Please ensure that for retainer jobs, that the Xflow AI is processed and ordered successfully and returning production files (STL and PTS files) before you proceed to sending jobs to the LAC. This is a critical step that is required for the LAC to process the order successfully.
To manufacture retainers, you will need to follow the following steps:
File requirements:
Xflow AI trim path file outputs
Aligner model requirements:
LAC requirements:
In case of an error at a customer, the respective error code is to be searched for. For each error, the cause and possible solutions are described. If the error cannot be solved with the available descriptions, it must be forwarded to Dental Axess Support.
Basically, the STL and PTS files of the affected jobs should be retrieved from the Xflow portal.
[10000] An Error Occurred
Error description: An undefined error has occurred. This can be due to a bug or an unhandled issue within the software.
Solution: Check STL and PTS Files.
[10001] Processing failed. The clamp ring must be closed during cutting.
Error description: During the processing of the job, a sensor indicated that the clamp ring was opened.
Solution: You may not have closed the clamping ring tightly enough before starting the job. The clamping ring must always be closed completely and, if necessary, also with some force.
[10002] Processing failed. The safety door must be closed during cutting.
Error description: While the job was being processed, a sensor indicated that the safety door had been opened.
Solution: You may not have closed the safety door tightly enough before starting the job. The safety door must always be closed completely and, if necessary, with some force. When closing the safety door, the handle should always be gripped centrally.
[10003] Processing failed. The suction unit must be turned on during cutting.
Error description: During the processing of the job, a sensor indicated that the suction was not detected or too little vacuum was generated by the suction.
Solution: Change the LAC Suction Unit filters if necessary. The LAC Suction Unit must not be switched off during laser processing of the LAC.
[10004] Image acquisition failed.
Error description: Capturing an image by the internal camera has failed.
Solution: Switch off the LAC and switch it back on after about one minute.
[20201] The data could not be prepared for cutting. Please check that the trim line matches the model.
Error description: The data could not be processed because the points of the PTS are too far from the surface of the STL.
Solution: Check STL and PTS Files.
[20401] The data could not be prepared for cutting. The data points are not within the boundaries.
Error description: The data led to a collision of the positioning system with the laser tip when calculating the cutting path.
Solution: Trimline and/or model must be recalculated. The model must comply with the given conditions.
[20501] The model data did not match with the detected aligner. Please make sure the correct aligner is inserted.
Error description: The internal camera could not match the aligner to be cut with the data from STL and PTS.
Solution: Check whether you have inserted the correct aligner. The aligner must also have been thermoformed with a model that corresponds exactly to the digital STL. After 3D printing the STL, the model must not be changed or adjusted manually.
We understand that there are other issues that can go wrong, so our team at Dental Axess is always available to assist you when things go wrong. Please feel free to reach out to us.